Experiment: TDK-Lambda Z+ PSU Integration and Handshake Validation

This document summarizes the integration and verification of the TDK-Lambda Z+ series power supply (Z+100-2) within the HAL. This experiment identified critical hardware and software configuration requirements for stable USB communication.

Hardware Setup

  • Instrument: TDK-Lambda Z+100-2 Power Supply.
  • Connection: USB (Serial-over-USB).
  • Interface: Internal FTDI Serial Bridge.

Essential Device Configuration

1. Front Panel Physical Setup (The "USB Mode" Fix)

By default, many Z+ units are configured for RS232/RS485 communication via the rear DB-sub connector. The USB port will appear dead unless the interface is explicitly switched to USB mode on the front panel.

[!IMPORTANT] Switching to USB Mode: 1. Press the [REM/LOC] button until the display shows the configuration menu. 2. Rotate the [VOLTAGE] encoder until you find the IF (Interface) menu. 3. Press the encoder to enter, and rotate until USB is displayed. 4. Press the encoder again to confirm. 5. Power Cycle: It is recommended to restart the unit after changing the interface mode.

2. SCPI Handshake (INST:NSEL)

Even in USB mode, the unit remains in a "dormant" SCPI state. - Mandatory Command: INST:NSEL 6 must be the first command sent. - Effect: This selects the specific instrument channel for the SCPI parser. Without this, the unit will not respond to *IDN?.

3. Communication Parameters

  • Baud Rate: 9600 (Must match the value set in the front panel BAUD menu).
  • Line Termination: Strictly \r\n (CRLF).

Results and Performance

The integration of these physical and software fixes resulted in a robust connection that survives instrument restarts.

Discovery Latency

Method Latency Status
Cold Scan (Full NI-VISA) ~10.6 seconds Pass
Cached Discovery (Nitro) ~0.45 seconds Pass

Verification Example

from instrumation.factory import get_instrument

# Ensure the PSU is set to USB mode on the front panel before running
psu = get_instrument("AUTO", "PSU")
print(f"Verified Connection: {psu.get_id()}")

psu.set_voltage(5.0)
psu.set_output(True)
psu.shutdown_safety()