Skip to content

Supported Instruments

This page lists all instrument models that Instrumation supports — both the specific models that have been validated and the broader families that share the same SCPI command set.

[!NOTE] Compatibility matrix convention (see issue #167): every driver section lists a Validated Model — the specific unit tested against real hardware — and an Also likely compatible list of sibling models that share the same SCPI command set but have not been verified in-lab. When one driver class is routed from multiple distinct IDN branches in factory.py (e.g. Keysight34461A serving both the Truevolt and legacy 34401-series families), each branch's models are called out explicitly with their own compatibility status, since a sub-model's SCPI dialect can diverge silently. Treat "assumed compatible" as best-effort: fall back to "GENERIC" passthrough if an untested model raises SCPI errors on a typed command.


Oscilloscopes

Keysight InfiniiVision

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightInfiniiVision DSOX2002A DSOX/MSOX 2000–6000 Series DSO-X, MSO-X, DSOX, MSOX

Also likely compatible (same InfiniiVision SCPI set): - DSOX1002A–DSOX1004A, DSOX1102G, DSOX1204G, DSOX2004A–DSOX2012A, DSOX3002T–DSOX3054T - MSOX1002A–MSOX1004A, MSOX2004A–MSOX3054T - DSOX4022A–DSOX4154A, MSOX4002A–MSOX4154A (4000 X-Series) - DSOX6002A–DSOX6054A, MSOX6002A–MSOX6054A - EDUX1052A/G

The 4000 X-Series (DSOX4xxxA/MSOX4xxxA) shares the same core :RUN/:STOP/:SINGLE, :WAVeform:* (WORD/LSBFirst preamble scaling) and :MEASure:* dialect this driver uses, so it auto-detects and runs through the generic DSOX/MSOX IDN keywords. Its segmented-memory acquisition is a capability this driver does not expose; treat the 4000 X-Series as assumed-compatible for the shared scope/measure API only.

Rigol DS1000Z / MSO1000Z

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDS1054Z DS1054Z DS1000Z/MSO1000Z Series DS1054Z, DS1074Z, DS1104Z, MSO1054Z, DS1000Z, MSO1000Z

Also likely compatible (same MSO1000Z/DS1000Z SCPI set): - DS1054Z, DS1074Z, DS1102Z, DS1104Z, DS1202Z, DS1204Z - MSO1054Z, MSO1074Z, MSO1102Z, MSO1104Z, MSO1202Z, MSO1204Z - DSEQ (DS1000Z-EDU variants)

[!NOTE] The driver implements universal DS1000Z SCPI commands only. LA (logic analyzer), :SOURce (AWG), :DECoder, :MASK, and :FUNCtion commands are reserved for the -S variants or paid options and are not covered.

Rigol MSO5000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolMSO5000 MSO5354 MSO5000 Series RIGOL + MSO5

Basic control (:RUN/:STOP/:SINGle/:AUTOscale) matches the MSO1000Z/DS1000Z family, but the measurement subsystem differs: :MEASure:ITEM? <item>,<src> takes the source as an explicit second argument on every query rather than DS1000Z's per-item :MEASure:FREQuency? CHANnel<n> form.

Also likely compatible (same MSO5000 command set): - MSO5072, MSO5074, MSO5102, MSO5104, MSO5204, MSO5354

[!NOTE] LA (logic analyzer), protocol decode, and mask-test features from the Programming Guide are not implemented here.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Siglent SDS

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSDS SDS1000 Series SDS Series SIGLENT

Also likely compatible: - SDS1002X-E, SDS1004X-E, SDS1102X-E, SDS1104X-E, SDS1202X-E, SDS1204X-E - SDS2002X, SDS2004X, SDS2102X, SDS2104X, SDS2202X, SDS2204X

Siglent SDS2000X Plus

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSDS2000XPlus SDS2102X Plus SDS Series (newer command set) SIGLENT + PLUS (with SDS1/SDS2/SDS5)

Uses the newer standard :SUBsystem:keyword command set from the SDS Series Programming Guide (EN11D+) -- :TRIGger:RUN/:TRIGger:STOP/ :TRIGger:MODE SINGle and :MEASure:SIMPle:VALue? <type> -- distinct from the legacy SiglentSDS driver's older ARM/TRSE/C<n>:PAVA? command shapes still used by the original SDS1000/SDS2000X(-E).

Also likely compatible (same newer command set): - SDS1000X HD, SDS5000X, SDS2000X Plus family models (any ...Plus/HD variant)

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Tektronix TDS

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
TektronixTDS TDS Series Tek TDS/TPS/MDO TEKTRONIX (non-AFG)

Rohde & Schwarz / Hameg HMO Compact

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RohdeSchwarzHMOCompact HMO722 HMO Compact Series HAMEG/ROHDE/ROHDE&SCHWARZ + HMO

Also likely compatible (same HMO Compact SCPI set): - HMO724, HMO1022, HMO1024, HMO1522, HMO1524, HMO2022, HMO2024

[!NOTE] Edge trigger only. Math functions (FFT, add/subtract/multiply), protocol decode, and mask-test features from the Programmer's Manual are not yet implemented.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Also likely compatible: - TDS1001C–TDS1012C, TDS2001C–TDS2024C, TDS2012B, TDS2022B - TPS2002B–TPS2024B - MDO3012–MDO3054 (basic scope commands)


Spectrum Analyzers

Keysight MXA / PXA

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightMXA MXA N9020A MXA Series MXA
KeysightPXA PXA N9030A PXA Series N9030, PXA

[!NOTE] KeysightPXA (v0.8.0+) ships 29 PXA-specific SCPI methods — measurement configuration, advanced triggering, enhanced markers, bandwidth/sweep, Real-Time Spectrum Analysis, and system queries — on top of the shared spectrum analyzer base API.

Also likely compatible: - MXA N9010A/B (EXA), N9020A/B, N9021A - PXA N9030A, N9040B

Rigol DSA

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDSA DSA800 Series DSA800 RIGOL (non-scope)

Also likely compatible: - DSA815, DSA815-TG, DSA832E, DSA875E

Rohde & Schwarz

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RohdeSchwarzSA FSV / FSW Series R&S SA ROHDE&SCHWARZ, R&S

Also likely compatible: - FSV, FSVR, FSW, FSWP, FPS, FPL

Anritsu

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
AnritsuSA MS2830A MS2830/MS269x ANRITSU

Also likely compatible: - MS2830A, MS2690A, MS2691A, MS2692A

Siglent SSA3000X

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSSA3000X SSA3021X SSA3000X Series SIGLENT + SSA

Standard SCPI-99 spectrum-analyzer subsystem, the same command shape as RigolDSA/AnritsuSA.

Also likely compatible: - SSA3011X, SSA3015X, SSA3032X, SSA3000X Plus, SSA3000X-R, SVA1000X

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Signal Generators

Keysight EXG / MXG

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightSG MXG N5183B EXG/MXG/PSG Series N5181, N5182, N5183, PSG, MXG, EXG

Also likely compatible: - EXG N5171B, N5172B, N5173B - MXG N5181A/B, N5182A/B, N5183A/B - PSG E8257D, E8267D

Rohde & Schwarz

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RohdeSchwarzSG SMA100B / SMC100A R&S SG ROHDE&SCHWARZ (SG context)

Also likely compatible: - SMA100A/B, SMB100A, SMC100A, SMF100A

Anritsu

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
AnritsuSG MG3700A MG3700 Series ANRITSU (SG context)

Tektronix AFG

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
TektronixAFG AFG3022C AFG3000 Series TEKTRONIX + AFG

Also likely compatible: - AFG3011C, AFG3021C, AFG3022C, AFG3102C, AFG3252C

Siglent SDG2000X

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSDG2000X SDG2042X SDG Series AWG SIGLENT + SDG

Uses Siglent's C<n>:BSWV (Basic Wave) command: a single keyword takes comma-separated PARAM,value pairs (C1:BSWV FRQ,1000, C1:BSWV AMP,1) rather than one SCPI subtree per parameter -- distinct from the Tektronix AFG3000's SOURce<n>:FREQuency:FIXed-per-parameter dialect. Has no frequency-list sweep mode, so configure_list_sweep() logs an unsupported-feature warning.

Also likely compatible (same BSWV/OUTP command set): - SDG1000X, SDG2000X Plus, SDG6000X

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rigol DG4000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDG4000 DG4062 DG1000Z/DG4000/DG5000 Series RIGOL + (DG4, DG5, DG1000Z)

Uses the Agilent/Keysight-33500-derived [:SOURce[<n>]]:APPLy:<shape> single-command style shared across the DG1000Z/DG4000/DG5000 families -- distinct from Tektronix AFG3000's subtree-per-parameter dialect and Siglent SDG's comma-separated BSWV keyword. Has no frequency-list sweep mode in this base command set, so configure_list_sweep() logs an unsupported-feature warning.

Also likely compatible (same APPLy/SOURce command set): - DG4102, DG4162, DG4202, DG1022Z, DG1032Z, DG5071, DG5102, DG5252

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

GW Instek MFG-2000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
GWInstekMFG2000 MFG-2120 MFG-2000 Series GW/GW INSTEK + MFG

Combines Rigol-style SOURce<n>:APPLy:<shape> waveform selection with a separate per-parameter SCPI subtree for frequency/amplitude/offset/ phase (SOURce<n>:FREQuency, SOURce<n>:AMPLitude, SOURce<n>:DCOffset, SOURce<n>:PHASe) -- unlike Rigol's single combined-argument APPLy call, MFG-2000 setpoints are each sent independently after selecting the waveform shape. Roadmap #176.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Stanford Research Systems DS345

Driver Validated Model Command Family Auto-Detect IDN Keywords
SRSDS345 DS345 SRS terse mnemonics (non-SCPI-subtree) STANFORD + DS345

Uses SRS's distinctively terse, colon-free command mnemonics (FREQ, AMPL, OFFS, FUNC, PHSE) rather than a SCPI :SOURce:* subtree; waveform shape is a numeric code (0=sine, 1=square, 2=triangle, 3=ramp, 4=noise, 5=arbitrary) instead of a mnemonic string. The DS345 has no software output-enable command -- the configured waveform is always live on the output BNC -- so set_output() logs an unsupported-feature warning and get_output() always reports True. Likewise there is no reference-clock command: set_reference_clock() logs an unsupported-feature warning (the rear-panel timebase input locks to an external 10 MHz on its own). Modulation is MTYP (type: 0=lin sweep, 1=log sweep, 2=AM, 3=FM, 4=PM, 5=burst) plus MENA (enable), and a frequency sweep is STFR/SPFR/RATE + MTYP 0 + MENA 1.

[!WARNING] Not yet verified against real hardware. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.


Network Analyzers

Keysight PNA

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightPNA PNA N5232A PNA / PNA-L / PNA-X E83, N52, PNA

Also likely compatible: - E8361C, E8362B, N5221A, N5222A, N5224A, N5225A, N5227A, N5230A–N5247A

Keysight FieldFox

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightFieldFox N9913A / N9914A FieldFox SA+VNA N99, FIELD FOX

Anritsu ShockLine / MS2035B

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
AnritsuShockLineVNA MS2035B ShockLine Series MS2035

Also likely compatible: - MS2034B, MS2035B, MS2036A, MS2037A, MS2038C, MS2039C, MS2047A, MS2060A

Siglent SNA5000A

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSNA5000A SNA5012A SNA5000A/X Series SIGLENT + SNA

Standard SCPI-99 network-analyzer subsystem, the same command shape as KeysightPNA -- Siglent's SNA5000A documentation is explicitly SCPI-99 compliant for this measurement class.

Also likely compatible: - SNA5014A, SNA5024A, SNA5044A, SNA5000X series

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Multimeters

Keysight 34461A Truevolt

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Keysight34461A 34461A Truevolt DMM 34461, 34460, 34401, 34410, 34411, 34420

Keysight34461A is shared across two IDN branches in factory.py (~line 423 and ~line 429, see issue #167): the Truevolt 34460/34461 family and the older 34401/34410/34411/34420 family. Only 34461A has been tested against real hardware; the rest are assumed compatible based on their published SCPI command references, not verified in-lab:

Model Family Status
34461A Truevolt ✅ Validated
34460A Truevolt Assumed compatible (6.5 digit, same SCPI set as 34461A)
34401A Legacy Assumed compatible (base MEAS:/CONF: subset only)
34410A Legacy Assumed compatible (6.5 digit)
34411A Legacy Assumed compatible (6.5 digit)
34420A Legacy Assumed compatible (μV/μΩ nanovolt commands untested)

[!WARNING] The 34401/34410/34411/34420 family predates Truevolt and may not support every SCPI command Keysight34461A sends (e.g. newer SENSe subsystem extensions). If you hit an SCPI error on one of these models, prefer "GENERIC" passthrough or file an issue with the failing command.

Fluke 8845A/8846A

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Fluke8846A 8846A Fluke 884x Series FLUKE + (8845, 8846)

Fluke8846A is not validated against real hardware yet; it targets the :MEAS:*?/:CONF:* SCPI subset documented in the Fluke 8845A/8846A Programmer's Manual, mirroring Keysight34461A's command shapes:

Model Status
8846A Assumed compatible (6.5 digit, same measurement subsystem as 8845A)
8845A Assumed compatible (5.5 digit, same command set as 8846A)

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Keithley 2000

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Keithley2000 2000 Keithley 2000 Series KEITHLEY + 2000

Siglent SDM3000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSDM3055 SDM3055 SDM3000 Series SIGLENT + SDM

Standard SCPI-99 MEASure/CONFigure multimeter subsystem, the same command shape as the Keysight 34401A family and Fluke 884x — DCV, ACV, DCI, ACI, 2W/4W resistance, frequency, period, capacitance, continuity, diode, and RTD/thermistor temperature. set_nplc() is SDM3000-specific.

Also likely compatible (same SDM series command set): - SDM3045X, SDM3065X, SDM3055A

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rigol DM3000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDM3068 DM3068 DM3000 Series RIGOL + DM3

Unlike the SCPI-99 CONFigure/MEASure? split used elsewhere in this library, the DM3000 dialect selects the active function with :FUNCtion:* and reads it back with a bare :MEASure:<FUNC>? query; :MEASure {AUTO|MANU} toggles ranging globally instead of a per-function RANG:AUTO. set_digits() (5/6/7/INC/DEC) is DM3000-specific.

Also likely compatible (same DM3000 command set): - DM3058, DM3058E, DM3062, DM3064

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Keithley DMM6500

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeithleyDMM6500 DMM6500 DMM6500 native SCPI KEITHLEY + (DMM6500, 6500)

Targets the DMM6500's native SCPI command set (function selection via :SENSe:FUNCtion "<FUNC>" then :READ?), not the SCPI2000/SCPI34401 compatibility personalities or TSP scripting. set_nplc(), set_auto_range() and set_offset_compensation() take an explicit SCPI function-name argument since ranging/NPLC live under :SENSe:<FUNC>:*, not a single global subsystem.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Power Supplies

TDK-Lambda Z+

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
TDKLambdaZPlus Z+100-2 Z+ Series (Serial/USB) TDK-LAMBDA, Z+

Also likely compatible: - Z+36-10, Z+50-2.5, Z+60-5, Z+100-2, Z+100-7, Z+150-4, Z+200-2.5, Z+360-1.7

[!IMPORTANT] Z+ units must be switched to USB mode on the front panel and wake up with INST:NSEL 6 before SCPI communication.

BK Precision 9130B

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
BKPrecision9130B 9130B 9130B Series (Triple Output) B&K/BK PRECISION + 9130

Not yet validated against real hardware. Three independent output channels are addressed via INST:NSEL {1\|2\|3}; every PowerSupply interface method accepts an optional channel= argument (defaults to channel 1) to select which output it targets before sending the plain VOLT/CURR/OUTP/MEAS:*? command. set_tracking_mode() is 9130B-specific and links CH2/CH3 outputs.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rigol DP800 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDP832 DP832 DP800 Series (Triple Output) RIGOL + DP8

Unlike the BK Precision 9130B's INST:NSEL-then-plain-command style, DP800 series commands address a channel directly via a [:SOURce<n>] numeric suffix or an explicit CH1|CH2|CH3 argument on :OUTPut/:MEASure — no channel-select round-trip required. Every PowerSupply interface method accepts an optional channel= (defaults to 1). set_tracking_mode() is DP800-specific and links CH2/CH3.

Also likely compatible (same DP800 command set): - DP831, DP831A, DP832A, DP821, DP821A, DP811, DP811A

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Siglent SPD3303X

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSPD3303X SPD3303X SPD3303X Series SIGLENT + SPD

CH1/CH2 are independently controlled outputs addressed with a CH1/ CH2 command prefix; CH3 is a fixed 5V/2.5A logic supply with output-enable control only (no settable voltage/current, and set_ovp/set_ocp/clear_protection are unsupported — the SPD3303X has no software OVP/OCP command). set_track_mode() selects independent/series/parallel CH1+CH2 coupling.

Also likely compatible: - SPD3303X-E, SPD3303C

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Keysight E36300 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightE36313A E36313A E36300 Series (Triple Output) KEYSIGHT/AGILENT + (E3631, E36313, E3633)

Uses the SCPI-99 channel-list syntax (@<chanlist>) shared with other Keysight instruments -- every setpoint/measurement command takes an explicit (@<n>) argument, distinct from the BK Precision 9130B's channel-select round-trip and the Rigol/Siglent CH<n> prefix style. set_output_pairing() is E36300-specific (OFF/series/parallel CH1+CH2).

Also likely compatible: - E36311A, E36312A

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Korad KA3005P

Driver Validated Model Command Family Auto-Detect IDN Keywords
KoradKA3005P KA3005P Korad flat ASCII (non-SCPI) KORAD

Not a SCPI instrument. Speaks Korad's own flat ASCII command set over serial (9600 8N1, no line termination, no error queue, no *RST/*CLS/*OPC?) -- this driver overrides check_errors, sync_config, and clear_status as no-ops and skips the SYST:ERR? round-trip in safe_send/query_ascii entirely. preset() emulates a reset by forcing output off, 0V, 0A limit since the hardware has no reset command. get_output() decodes bit 6 of the single-byte STATUS? response. Widely cloned by Tenma, RS, Velleman, and Stamos under different model numbers with an identical protocol.

[!WARNING] Not yet verified against real hardware. If you hit a protocol error, prefer "GENERIC" passthrough or file an issue with the failing command.

GW Instek GPP Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
GWInstekGPP4323 GPP-4323 GPP Series (4 independent channels) GW/GW INSTEK + GPP

The GPP dialect puts the channel number directly as a numeric suffix on the keyword itself (SOURce2:VOLTage, OUTPut2:STATe, MEASure2:CURRent?), distinct from every other multi-channel PSU dialect already in this library (Keysight (@<n>), Rigol/Siglent CH<n>: prefix, BK Precision channel-select round-trip). Channel 1's suffix may be omitted on real hardware but this driver always sends it explicitly for clarity. set_series_mode()/set_parallel_mode() link CH1+CH2 (and CH3+CH4 where present).

Also likely compatible (same GPP command set): - GPP-1326, GPP-2323, GPP-3323

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Agilent/HP 6632B/6634B Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Agilent6632B 6632B 663xB System DC Power Supply KEYSIGHT/AGILENT/HEWLETT-PACKARD/HP + (6631, 6632, 6633, 6634)

Legacy single-channel SCPI-99 PSU family, routed through the same Keysight/Agilent IDN branch as Keysight34461A/KeysightPNA since it predates the Keysight brand name but shares the same SCPI lineage. set_autostart() maps to OUTP:PON:STAT {RCL0|RST} (power-on state).

Also likely compatible: - 6631B, 6633B (same programming guide, different voltage/current ranges)

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

AIM-TTi CPX400DP

Driver Validated Model Command Family Auto-Detect IDN Keywords
AimTTiCPX400DP CPX400DP AIM-TTi PowerFlex command set AIM-TTI/THURLBY THANDAR/AIMTTI + CPX400

Puts the channel number directly on the command keyword with no colon or prefix separator (V1 <v>, I2 <i>, OP1 {0|1}) -- distinct from every SCPI-99 SOURce<n>:-prefixed or INST:NSEL-selected dialect elsewhere in this library. Measured readbacks (V<n>O?/I<n>O?) may echo a trailing unit suffix ("4.999V"); the driver strips it before parsing. set_track_mode() maps to CONFIG {0|1|2} (independent/ series/parallel).

[!WARNING] Not yet verified against real hardware. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rohde & Schwarz HMP4040 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RohdeSchwarzHMP4040 HMP4040 HMP Series (4 independent channels) HAMEG/ROHDE/ROHDE&SCHWARZ + HMP

Channels are selected via INST:NSEL {1..4} and then addressed with plain VOLT/CURR/OUTP commands -- the same select-then-command shape as BKPrecision9130B, distinct from the GW Instek GPP's numeric-suffix-on-keyword dialect. set_global_output() maps to OUTP:GEN {ON|OFF}, the master enable for channels armed via OUTP:SEL (not wired up individually here). clear_protection() disables the fuse-link coupling (FUSE:STAT OFF) since the HMP series has no separate OVP/OCP trip-clear command.

Also likely compatible (same SCPI dialect, fewer channels): - HMP2020, HMP2030, HMP4030

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Ametek Sorensen SG Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SorensenSG SG Series SG/SGA/SGX Series (core PSU subsystem) SORENSEN

Standard SCPI-99 :SOURce/:OUTPut/:MEASure subsystem, directly comparable to KeysightE36313A/RigolDP832. The Sorensen-specific :PROGram subsystem for stored multi-step voltage/current sequences is intentionally not implemented in this initial driver -- it is a genuinely new interaction pattern (triggered sequence playback) not present in any other PowerSupply driver in this library, and is left for a future pass.

Also likely compatible: - SGA, SGX series (same core SCPI subsystem; :PROGram support varies)

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

BK Precision 1685B/1687B/1688B Series

Driver Validated Model Command Family Auto-Detect IDN Keywords
BKPrecision1685B 1685B BK Precision flat ASCII (non-SCPI) B&K/BK PRECISION + (1685, 1687, 1688)

Not a SCPI instrument. Speaks the same flat ASCII command family as KoradKA3005P (VSET1:<v>, ISET1:<i>, OUT1/OUT0) over a USB virtual COM port -- distinct from the SCPI-based BKPrecision9130B driver, which targets BK Precision's other, multi-channel product line. check_errors, sync_config, and clear_status are no-ops; preset() emulates a reset (no hardware *RST) by forcing output off, 0V, 0A limit.

Also likely compatible: - 1687B, 1688B (same programming manual, different voltage/current ranges)

[!WARNING] Not yet verified against real hardware. If you hit a protocol error, prefer "GENERIC" passthrough or file an issue with the failing command.


AC Power Sources

Keysight AC6800B Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightAC6800B AC6800B Series AC6800B Series Programming Guide KEYSIGHT + AC68

New ACPowerSource base class (src/instrumation/drivers/base.py) -- every PowerSupply driver in this library is DC-only (a single voltage/current setpoint with no frequency/phase/output-mode concept), which does not fit an AC source needing frequency programming and an explicit AC/DC/AC+DC output mode. Covers voltage/frequency setpoints (SOURce:VOLTage/:FREQuency), output mode select (SOURce:VOLTage:MODE {AC|DC|AC+DC}), output on/off, true-RMS voltage/current/power measurement (MEASure:VOLTage:AC?/ :CURRent:AC?/:POWer:AC?), OVP/OCP, and DC offset (for AC+DC mode). Registered under a new "ACPSU" driver type to avoid conflating it with the DC-only "PSU" category.

Per issue #192's scope, transient/waveform-step programming and arbitrary-waveform output are not implemented in this initial driver -- left as a follow-up, consistent with how SorensenSG deferred its :PROGram subsystem.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Electronic Loads

Siglent SDL1000X

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
SiglentSDL1000X SDL1000X SDL1000X Series SIGLENT + LOAD

Also likely compatible: - SDL1020X, SDL1030X, SDL1060X

BK Precision 8600 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
BKPrecision8600 8600 8600 Series B&K/BK PRECISION + (8600, 8601, 8602, 8610, 8612, 8614, 8620)

Not yet validated against real hardware. Supports CC/CV/CR/CP modes, OVP/OCP/OPP protection, and BK-specific battery discharge test mode (set_battery_test_mode, set_battery_cutoff_voltage, get_battery_test_capacity). List-mode and transient-stepping programming from the 8600 Series manual are not yet implemented.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rigol DL3000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RigolDL3021 DL3021 DL3000 Series RIGOL + DL3

Supports CC/CV/CR/CP modes and battery discharge test mode (set_battery_test_mode, get_battery_test_capacity, get_watt_hours, get_discharging_time). Unlike the BK Precision 8600, the DL3000 series has no software OVP/OCP/OPP trip-point subsystem in its base command set — front-panel OCP/OPP are test modes, not simple protection levels — so set_ovp/set_ocp/ set_opp/clear_protection log an unsupported-feature warning here.

Also likely compatible (same DL3000 command set): - DL3031, DL3031A

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

ITECH IT8500+ Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
ItechIT8512Plus IT8512+ IT8500+ Series ITECH + IT85

Supports CC/CV/CR/CP modes plus CURRent:PROTection/POWer:PROTection (OCP/OPP trip levels) and a simulated short-circuit input (set_short_circuit). The IT8500+ series has no dedicated software OVP trip-point command in its base command set, so set_ovp/ clear_protection log an unsupported-feature warning here.

Also likely compatible (same IT8500+ command set): - IT8511+, IT8513+, IT8511A+, IT8512A+, IT8513A+

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Chroma 63200A Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Chroma63200A 63200A series 63200A Series (High Power) CHROMA + 632

The most distinctive load dialect in this library: a single MODE mnemonic encodes BOTH the operating mode and measurement range (MODE CCH = CC High range, MODE CVL = CV Low range, etc.). set_mode() always selects the High-range variant as the safest default for an unknown DUT — send write("MODE CCL")/"MODE CCM" directly for Low/Middle range. Setpoints use the <FUNC>:STATic:L1 level-1 static value (the front panel's "A" state); the L2 ("B" state) and slew-rate parameters are not wired up. No dedicated software OVP/OCP/OPP trip-level commands exist — OCP/OPP are selectable MODEs, not settable protection levels — and there is no direct MEASure:RESistance? query, so those methods log an unsupported-feature warning.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Prodigit 3311F

Driver Validated Model Command Family Auto-Detect IDN Keywords
Prodigit3311F 3311F 3310F Series plug-in module PRODIGIT + (3311, 3310)

The 3311F is a plug-in module hosted in a 3302F mainframe chassis, not a standalone load -- every command is prefixed with a CHAN<n> slot select (default slot 1, override via slot= on any method), an extra addressing layer beyond standalone loads like SiglentSDL1000X or ItechIT8512Plus. The CC/CV/CR/CP command shape itself (MODE:CC/CURR <a>/LOAD ON/MEAS:VOLT?) is otherwise directly comparable.

[!WARNING] Not yet verified against real hardware. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.


Lock-In Amplifiers

New instrument category (issue #204). LockInAmplifier is a new abstract base in drivers/base.py; "LOCKIN" is a new canonical driver_type key registered in factory.KNOWN_DRIVER_TYPES, with a SimulatedLockInAmplifier registered for SIM mode.

Stanford Research Systems SR830

Driver Validated Model Command Family Auto-Detect IDN Keywords
SRSSR830 SR830 SRS terse mnemonics (non-SCPI-subtree) STANFORD + SR830

Like the SRS DS345, uses terse colon-free command mnemonics (FREQ, PHAS, SLVL, SENS, OFLT, HARM) rather than a :SOURce:*/:SENSe:* subtree. Sensitivity and time-constant values are indexed by integer code, not physical units directly -- set_sensitivity()/set_time_constant() accept a physical value and round it to the nearest supported code. measure_xy()/ measure_r_theta() use SNAP? to read multiple parameters at a single instant, avoiding the time skew that separate OUTP? queries would introduce for a fast time constant. Roadmap #180.

[!WARNING] Not yet verified against real hardware. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.


LCR Meters

New instrument category (issue #204). LCRMeter is a new abstract base in drivers/base.py; "LCR" is a new canonical driver_type key registered in factory.KNOWN_DRIVER_TYPES, with a SimulatedLCRMeter registered for SIM mode.

Keysight E4980A/AL

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightE4980A E4980A E4980A/AL Precision LCR Meter KEYSIGHT/AGILENT + E4980

set_measurement_function() selects the impedance parameter type (CPD, CPQ, CSD, LSD, RX, ZTD, etc.); measure() returns a (primary, secondary) tuple, e.g. capacitance and dissipation factor for CPD. set_bias_voltage/set_bias_state control the DC bias source.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Hioki IM3536

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
HiokiIM3536 IM3536 IM3523/IM3533/IM3536/IM3570/IM3590 family HIOKI + IM35

SCPI-style command set shared across the IM35xx/IM3570/IM3590 family. :FREQuency/:FREQuency? confirmed against Hioki's own PLC-integration examples; parameter-type selection and measurement readback follow the same :FUNC:IMP/:MEASure? shape as the Keysight E4980A LCR meter class.

[!WARNING] Best-effort: the full IM3536 command reference requires a Hioki-issued LCR Application Disc not published as a standalone PDF. Not yet verified against real hardware -- prefer "GENERIC" passthrough or file an issue with the failing command.

Keysight E4990A Impedance Analyzer

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightE4990A E4990A E4990A SCPI Command Reference KEYSIGHT/AGILENT + E4990

Unlike the fixed-frequency KeysightE4980A/HiokiIM3536 LCR meters, the E4990A sweeps frequency (1 MHz-120 MHz) and returns trace data -- closer in shape to the KeysightPNA VNA driver, but measuring impedance/LCR parameters (CALCulate:PARameter1:DEFine) instead of S-parameters. LCRMeter.measure() performs a single-point measurement at the current CW frequency (SENS:FREQ:CW); get_trace_data()/ get_complex_trace() fetch a full swept trace in formatted (primary, secondary) or raw complex (real, imaginary) form (CALC:DATA:FDATa?/:SDATa?). Sweep configuration (set_start_frequency/set_stop_frequency/set_points/ set_sweep_type) mirrors the existing NetworkAnalyzer sweep API. connect() sets the safety-guard frequency range to the instrument's 1 MHz-120 MHz span.

Per issue #193, equivalent-circuit fitting and fixture compensation (open/short/load cal) are out of scope for this initial driver and are left as a follow-up.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Frequency Counters

Keysight 53230A

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Keysight53230A 53230A 53230A/53220A Universal Counter KEYSIGHT/AGILENT/HP + (53230, 53220, 53181)

Also likely compatible: - 53220A (350 MHz), 53230A (350 MHz, 12 digits/s), 53181A (225 MHz, RF counter predecessor)

[!NOTE] Prior to issue #204, this driver had no explicit IDN branch in factory.py and relied on the single-registered-candidate fallback; now that a second "COUNTER" driver (FlukePM6690) is registered, an explicit KEYSIGHT/AGILENT/HP + model-number branch is required.

Fluke PM6690

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
FlukePM6690 PM6690 PM6690 native SCPI FLUKE + PM6690

The PM6690 has two GPIB personalities: a native SCPI mode (targeted here) with a command set optimized for its own capabilities, and a compatible mode that emulates a Keysight/HP 53131A/53132A. Native mode uses the same standard SCPI-99 counter subsystem shape as Keysight53230A.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Data Acquisition Units

Keysight DAQ970A/DAQ973A

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightDAQ970A DAQ970A DAQ970A/DAQ973A Programming Guide KEYSIGHT/AGILENT + (DAQ970, DAQ973)

New DataAcquisitionUnit base class (src/instrumation/drivers/base.py), registered under a new "DAQ" driver type -- the direct, backward-compatible successor to the 34970A/34972A Data Acquisition/Switch Unit. Unlike a Multimeter, a DAQ/switch unit multiplexes many channels through plug-in modules behind a single measurement engine, addressed with SCPI channel-list syntax (slot is the hundreds digit, channel the tens/units -- e.g. (@101,102,203)). DataAcquisitionUnit.format_channel_list() builds this syntax from a plain list ([101, 102, 203]) or a range string ("101:110").

Covers per-channel measurement-function configuration (CONFigure:VOLTage:DC/:AC, :RESistance/:FRESistance 2W/4W, :TEMPerature, :FREQuency), scan-list setup and triggering (ROUTe:SCAN, INITiate, TRIGger:SOURce, FETCh?), immediate per-channel reads (READ?), and relay open/close for switch modules (ROUTe:CLOSe/:OPEN/:CLOSe?).

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


USB Power Sensors

New instrument category (issue #204). "SENSOR" is a new canonical driver_type key registered in factory.KNOWN_DRIVER_TYPES.

Keysight U2000 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
KeysightU2000 U2004A U2000 Series USB Power Sensor KEYSIGHT/AGILENT/HP + (U2000, U200)

Headless USBTMC-only sensor (no front panel); standard SCPI power- sensor command set (FETCh?/MEASure?, SENSe:FREQuency, SENSe:CORRection:GAIN2 for external gain/loss offset, CALibration:ZERO). set_frequency() sets the CW frequency used for the sensor's internal cal-factor table lookup, not an output.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Rohde & Schwarz NRP-Z Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
RohdeSchwarzNRPZ NRP-Z21 NRP-Z Series USB Power Sensor HAMEG/ROHDE/ROHDE&SCHWARZ + NRP

Headless USB sensor (no front panel), similar in shape to the Keysight U2000 series but with R&S-specific correction subsystems: SENSe:CORRection:DCYCle for pulse-power duty-cycle correction and SENSe:CORRection:SPDevice for S-parameter device correction.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


RF Switches

New instrument category (issue #204). "SWITCH" is a new canonical driver_type key registered in factory.KNOWN_DRIVER_TYPES.

Mini-Circuits RC/RFS Series

Driver Purpose Auto-Detect IDN Keywords
MiniCircuitsRCSwitch RF switch matrix control (RC-4SPDT, RC-1SP4T, RC-2SP4T, etc.) MINI-CIRCUITS/MINICIRCUITS

Not a SCPI instrument. Uses Mini-Circuits' own flat ASCII command set over USB/Ethernet: SET{switch}={0|1} for individual SPDT switches (A-H), SP{n}T:STATE:PORT {port} for SPnT port selection. Roadmap #174.

[!WARNING] Could not be verified against Mini-Circuits' official Programming Manual PDF at write time (host was unreachable from this environment). Confirmed only via third-party summaries of the SET[switch]=[state] command. Not yet verified against real hardware -- prefer "GENERIC" passthrough or file an issue with the failing command.


RF Peak Power Meters

Boonton 4530 Series

Driver Validated Model SCPI Family Auto-Detect IDN Keywords
Boonton4531 4531 (single-channel) SCPI 1993 BOONTON + (4531, 4530)
Boonton4532 4532 (dual-channel) SCPI 1993 BOONTON + 4532

New PowerMeter base class (src/instrumation/drivers/base.py) -- distinct from the POWERMETER-keyed TektronixPA1000/YokogawaWT310 AC power analyzers (registered separately, no shared ABC), this covers wideband RF peak/pulse power measurement: CW average power (MEAS:POWER?), peak/pulse power (MEAS:PEAK?), video bandwidth for pulse demodulation (SENS:BAND:VIDEO), trigger source/level (TRIG:SOURCE/TRIG:LEVEL), pulse-width measurement (MEAS:PULSE:WIDTH?), relative gain/loss offset (CAL1:OFFSET), and sensor zero calibration (CAL:ZERO). Registered under a dedicated "PEAKPM" driver type to avoid ambiguity with the existing "POWERMETER" category.

Boonton4532 (dual-channel) overrides measure_power/ measure_peak_power with an optional channel argument (1 or 2); Boonton4531 (single-channel) omits the parameter entirely. GPIB and RS-232 only -- no USB/LAN on this generation, which fits the existing PyVISA RealDriver path directly.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.


Power Analyzers

New instrument category (issue #204). "POWERMETER" is a new canonical driver_type key registered in factory.KNOWN_DRIVER_TYPES.

Yokogawa WT310 Series

Driver Validated Model Command Family Auto-Detect IDN Keywords
YokogawaWT310 WT310 WT310/WT310HC/WT332/WT333 YOKOGAWA + WT3

Uses Yokogawa's :NUMeric[:NORMal]:ITEM<x> output-item configuration model: each numeric output slot (1-255) is assigned a function (U=voltage, I=current, P=active power, S=apparent, Q=reactive, LAMBda=power factor) and element (1-3 for multi-element models), then :NUMeric[:NORMal]:VALue? reads back all configured items as a single comma-separated response -- distinct from a per-quantity MEASure:<FUNC>? query used by DMM-style instruments. measure_voltage/measure_current/measure_active_power/ measure_power_factor are convenience wrappers that configure slot 1 and read it back in one call.

[!WARNING] Not yet verified against real hardware. If you hit an SCPI error, prefer "GENERIC" passthrough or file an issue with the failing command.

Tektronix PA1000

Driver Validated Model Command Family Auto-Detect IDN Keywords
TektronixPA1000 PA1000 Community-verified quirky SCPI TEKTRONIX + PA1000

Tektronix's published PA1000 manual omits the SCPI programming chapter; this driver is built from community-verified, hands-on PyVISA usage notes and documents several real quirks:

  • Every command requires the leading : root designator.
  • Semicolon-separated command chaining is not supported.
  • No *OPC? support -- wait_ready() sleeps a fixed 0.5s settle time instead of polling; *RST needs a fixed 5s delay.
  • Uses :SEL:* to choose which readings appear in :FRD? output, then :FRD? fetches them as a comma-separated list -- clear_selection()
  • select_*() + fetch_readings() compose into the measure_voltage_rms/measure_current_rms/measure_active_power/ measure_frequency convenience wrappers.
  • Non-LXI-compliant Ethernet: raw socket on port 5025, no VXI-11.

[!WARNING] Not yet verified against real hardware by this project. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.


Temperature Controllers

Lake Shore Model 336

Driver Validated Model Command Family Auto-Detect IDN Keywords
LakeShore336 336 Model 336/335 ASCII mnemonic set LSCI/LAKE SHORE/LAKESHORE + (336, 335)

New TemperatureController base class (src/instrumation/drivers/base.py) -- a cryogenic temperature controller reads multiple sensor inputs (lettered A-D) and drives PID heater-output control loops (numbered 1-2), which has no equivalent in the DC-only PowerSupply ABC. Covers Kelvin/Celsius/sensor-units readings (KRDG?/CRDG?/SRDG?), per-loop setpoint and PID gain configuration (SETP/PID), heater range and output queries (RANGE/HTR?), ramp-rate warm-up/cool-down limiting (RAMP), control-mode selection (OUTMODE), and autotune (ATUNE). Alarm status (ALARMST?) is also exposed. Sensor input curve/type configuration (INTYPE) is passed through as a raw comma-separated parameter string, since the exact field encoding varies by sensor family (diode, RTD, thermocouple) and firmware revision.

Also likely compatible (same ASCII mnemonic command set): - Model 335 (2 inputs / 1 loop -- use loop 1 only)

[!WARNING] Not yet verified against real hardware. If you hit a command error, prefer "GENERIC" passthrough or file an issue with the failing command.


GENERIC (Universal Fallback)

Driver Purpose Auto-Detect IDN Keywords
GenericDriver Catch-all driver for unidentified instruments; accepts any model none (explicit "GENERIC" request or unrecognized *IDN?)

"GENERIC" is a first-class, always-registered key in DriverRegistry — get_drivers_by_type("GENERIC") always returns GenericDriver (real mode) and SimulatedGeneric (SIM mode). It is used when connect_instrument() cannot identify an instrument, or when you explicitly pass driver_type="GENERIC" to get_instrument(). It exposes only generic SCPI passthrough (no typed measurement API), so it is a safe default, never a silent DMM/SA misread — see issue #148.


Summary

Category Drivers Validated Models
Oscilloscopes 7 DSOX2002A, DS1054Z, SDS Series, TDS Series, HMO722, SDS2102X Plus, MSO5354
Spectrum Analyzers 5 MXA N9020A, PXA N9030A, DSA800, MS2830A, SSA3021X
Signal Generators 9 N5183B, AFG3022C, SMA100B, MG3700A, SMA100A, SDG2042X, DG4062, MFG-2120, DS345
Network Analyzers 4 N5232A, N9913A, MS2035B, SNA5012A
Multimeters 6 34461A, 2000, 8846A, SDM3055, DM3068, DMM6500
Power Supplies 12 Z+100-2, 9130B, DP832, SPD3303X, E36313A, KA3005P, GPP-4323, 6632B, CPX400DP, HMP4040, SG Series, 1685B
AC Power Sources 1 AC6800B Series
Electronic Loads 6 SDL1000X, 8600, DL3021, IT8512+, 63200A, 3311F
Lock-In Amplifiers 1 SR830
LCR Meters 3 E4980A, IM3536, E4990A
Frequency Counters 2 53230A, PM6690
Data Acquisition Units 1 DAQ970A
RF Switches 1 RC-4SPDT-A18
USB Power Sensors 2 U2004A, NRP-Z21
Power Analyzers 2 WT310, PA1000
RF Peak Power Meters 2 4531, 4532
Temperature Controllers 1 Model 336
Total 61